OmniFluo-900 Steady-State & Time-Resolved Fluorescence Spectrometer

Description

Steady-State & Time-Resolved Fluorescence Spectrometer

OmniFluo Fluorescence Systems adopt modular design, which absorb Zolix fifteen years design and manufacture experience of spectroscopy system. The open architecture design also allows for application changes. As your application needs to grow, so can your OmniFluo. By different accessories, OmniFluo also could realize PL, Raman, transmission, reflection, absorption and detector calibration measurement. The possible configurations are endless.

OmniFluo Fluorescence Systems adopt high performance Omni-λmonochromator/spectrograph, high sensitivity single or multi-channel detector, single photon counting/lock-in amplifier technique. The standard for sensitivity is the signal-to-noise ratio calculated from a water Raman spectrum. Using this standardized test, our signal-to-noise ratio is 12000:1.

OmniFluo Fluorescence Systems can combine the PL and PLE in the same architecture. There are many configurations to fit your specific measurement, such as fluorescence spectroscopy, excitation spectroscopy, normal and low temperature testing.

Benefits

•Modular design, different function perfect combination to fit your specific needs
•Extended wavelength range with triple grating monochromators, 200 nm-2500 nm*
•Emission spectroscopy correction function*
•High output power, 75W Xenon light source
•Different Laser wavelength option*, 375 nm/405 nm/442 nm/532 nm/785 nm/1064 nm.
•Infrared test functions – Expansion option*
•Quantum Yield – Expansion option*
•Polarization fluorescence – Extension option*
•Fluorescence lifetime — μs-s fluorescence lifetime measurement*
* Confirm the configurations according to the actual need.
Specifications
OmniFluo-960OmniFluo-990
Sensitivity (steady state)>12000:1>12000:1
Lifetime RangeSteady stateμs-s
Spectral Range200-900 nm
Extensible to infrared
200-900 nm
Extensible to infrared
MonochromatorsTypeC-TC-T
Focal length320 mm320 mm
Stray light rejection1*10^-51*10^-5
Resolution0.08 nm0.08 nm
Wavelength Accuracy± 0.2 nm± 0.2 nm
Wavelength Repeatability± 0.1± 0.1
Excitation SourcesType75W Xenon light source75W Xenon light source & 60W Microsecond flashlamp
ModeSteady stateSteady state & Time-Resolved
Pulse WidthN.A.2.9 μs
DetectorsPMTPMT
Data AcquisitionSingle photo countingSingle photo counting
Number of detector channels11
Min. width of time binsN.A.64ps
Computer interfaceUSBUSB
SoftwareSteady stateSteady state & Time-Resolved

 

Options

ModelDescription
OmniFluo-B1Infrared measurement accessories. Contains the corresponding detector and data acquisition, which can meet the 800-1700nm spectral range test.
OmniFluo-B2Infrared measurement accessories. Contains the corresponding detector and data acquisition, which can meet the 800-2500nm spectral range test.
OmniFluo-C1Quantum Yield measurement accessories.
OmniFluo-C2Polarization fluorescence measurement accessories.